Lear Siegler Australasia

Ambient Air QualityAmbient Air QualityAmbient Air QualityAmbient Air Quality

Model 5012 Multi Angle Absorption Photometer

Model 5012 Multi Angle Absorption Photometer Click here to enlarge image
Click here for Product Data Sheet

PRODUCT DESCRIPTION

The Moldel 5012 Multi Angle Absorption Photometer is a precision instrument for the measurement of black carbon in the atmosphere. Black carbon (BC), a product of incomplete combustion, comes from industrial pollustion, traffic, fires, the burning of coal and biomass fuels.

Unlike carbon dioxide emissions, which add to global warming by trapping heat in the atmosphere, soot emissions may contribute to global warming and climate change by absorbing sunlight, heating the air and making the atmosphere more unstable.

The Model 5012 MAAP measures atmospheric BC loading using a radiative transfer scheme to particle loaded glass fiber filters. The black carbon content of the aerosols are continuously determined by simultaneously measuring the optical absorption and scattering of ligh byh the particles collected on the filter tape.

The combination of these two techniques provides a much truer measurement of the black carbon content.

PRODUCT SPECIFICATIONS

Method

Multi Angle Absorption Photometer (MAAP)

Minimum Detection Limit

2 minute average <100ng/m3 BC             <0.66 M/m Babs

10 minute average < 50ng/m3 BC            <0.33 M/m Babs

30 minute average < 20ng/m3 BC            <0.13 M/m Babs

Filter Change

At 20% transmission (approximately 30µg

Active Measurement Time

>98%

Air Flow Regulation

Deviation <1% (PID type)

Filter Tape

Glass Fiber, type GF 10, approximately 40m

Data Output

2 serial interfaces RS232

Analog output : 0/4mA - 20mA or 0V - 10V

Data Memory

18,560 concentration values, log book with 1,632 entries, 385 daily averages

Operating Temperature

-4 to 122ºF (-20 to 50ºC)

Dimensions

Instrument: 19" (W) x 12.25" (H) x 13" (D) / 483mm (W) x 311mm (H) x 330mm (D)

Pump: 8.25" (W) x 8.75" (H) x 4.25" (D) / 210mm (W) x 222mm (H) x 108mm (D)
 

Power Supply

Instrument: 100-240V, 50/60 Hz, 330W max., 15W without pump

Pump: 100-110/100-120V, 50/60 Hz or 220/240V, 50/60 Hz, 100W

Pump Type

RPM regulated rotary vane pump

Weight

Instrument: 50lbs (22.5 kg)

Pump: 13.5lbs (6.1 kg)

PM10, PM2.5, PM1 and TSP inlets

Sampling Tube Extensions

Analog I/O Expansion Board

Weather Enclosure

Rack Mounting Accessories

Calibration Kits for Flow Rate, Temperature and Pressure Sensors

LSA Sales and Support Contact Us Today!

HEAD OFFICE
Unit 5A, 2 Resolution Drive
Caringbah, NSW 2229
Australia
Ph: 61 2 9531 5444
Fax: 61 2 9531 5411
sales@learsiegler.com.au