Detector
|
Chemiluminescence (CLD) Photodiode (thermally stabilized with Peltier Cooler)
|
NO/NOx Ranges
|
0- 3 to 0 - 3,000 ppm NO/NOx (Other Ranges Available) (Four user programmable ranges)
|
Response Time
|
Typically < 2 seconds to 90% Full Scale
|
Repeatability
|
Better than 0.5% of Full Scale
|
Linearity
|
Better than 0.5% of Full Scale
|
Noise
|
Typically less than 1% of Full Scale
|
Zero & Span Drift
|
Less than 1% of Full Scale per 24 Hours
|
Zero & Span Adjustment
|
Via front panel, TCP/IP or RS-232
|
NH3, HCN & SO2 Effect
|
Not detectable with 100 ppm
|
CO2 & H2O Effect
|
Less than 2.0% with 10% CO2 and Chiller
|
Flow Control
|
Electronic Proportional Pressure Controller
|
Sample Flow Rate
|
Typically 2.0 LPM (0.6 LPM with low flow option)
|
NO2 Converter
|
Vitreous Carbon Material @ 205°C >95% efficiency
|
Ozonator
|
Ultraviolet Lamp
|
Air or O2 Requirements
|
Less than 0.01 ppm NOX at 350 cc/Min. @ 25 psig (Dew Point < -10ºC)
|
NO/NOx Control
|
Manual/Remote/Auto Cycle
|
Outputs Available
|
TCP/IP, RS232, 0-1, 0-5, 0-10, VDC, 4-20mA (selectable)
|
Discrete Alarms/Control
|
15 definable, optically isolated solid state relays (60 VAC max @ 600mA max)
|
Digital Diagnostics
|
Control Voltages, Temperatures, Pressures, Flow Parameters
|
Keypad Displays
|
Factory Settings, TCP/IP Address, Passwords (4), Scalable Analog Output Voltages, Full Scale Range Select, Auto Cal Times
|
Special Features
|
Calculated NO2 derived from NOx converter efficiency, Auto Ranging, Auto Calibration (adjustable through internal clock). Less than 3 cc Gold Plated Reaction Chamber, Data Archiving
|
Display
|
3” x 5” Back lit LCD
|
Sample Temperature
|
Up to 100°C Non-condensing
|
Oven Temperature
|
85°C standard; up to 100°C on request
|
Ambient Temperature
|
5 to 40°C
|
Ambient Humidity
|
Less than 90% RH Non-condensing
|
Warm-Up Time
|
1 Hour (Typical)
|
Fittings
|
1/4 Inch Tube
|
Power Requirements
|
115 VAC/60Hz or 230 VAC/50 Hz;
|
Dimensions
|
5¼ H x 19 W x 23 D (Inches)
|
Weight
|
65 lbs.
|