Model 600L Nondispersive Infrared/Oxygen Analyser
|
PRODUCT DESCRIPTION
The Model 600L Series digital analyser by California Analytical Instruments optical bench consists of the IR source, sample cell and patented detector. The light source is electronically chopped and replaces the need for mechanical chopper there by eliminating all moving parts from the optical bench.
This product line is designed around a state of the art 16 bit microprocessor, with 16 digital inputs, 16 digital outputs, 16 analogue inputs and 4 analogue outputs. The analyser can be manually operated from the keypad or remotely via TCP/IP, RS-232C communications and discrete inputs.
The analyser display includes screen presentation of all analyser alarms and four levels of password protection are provided.
|
PRODUCT SPECIFICATIONS
|
IR Analysis Method
|
Non dispersive Infrared (NDIR)
|
NDIR Components
|
CO / CO2 / CH4 / C3H8 / N2O
|
Detector Type
|
Microflow
|
NDIR Ranges
|
From 0-50 ppm up to 0-100% (Dependent upon gas & cell length)
|
Range Ratio
|
10:1
|
Response Time (IR)
|
T90 <15Seconds to 60 Seconds Adjustable (Depending on configuration)
|
IR Sample Cell
|
Gold Coated Pyrex Glass
|
Resolution
|
Displays 1 or 2 Significant Digits
|
Repeatability
|
Better than 1.0% of Full Scale
|
Linearity
|
Better than 1% Full Scale
|
Noise
|
Less than .5% of Full scale
|
Zero & Span Drift
|
Less than 1% of Full Scale per 24 Hours
|
Zero & Span Adjustment
|
Via front panel
|
Sample Flow Rate
|
Typically 0.2 to 2.0 LPM
|
Oxygen Analysis Method
|
Paramagnetic or Electrochemical
|
O2 Ranges
|
0-25 to 0-100% O2 Full Scale
|
O2 Response Time
|
T90 < 10-25 Seconds
|
|
|
HEAD OFFICE
Unit 5A, 2 Resolution Drive
Caringbah, NSW 2229
Australia
Ph: 61 2 9531 5444
Fax: 61 2 9531 5411
sales@learsiegler.com.au