Pour Point Process Analyser

Pour Point Process Analyser Click here to enlarge image
Click here for Product Data Sheet

PRODUCT DESCRIPTION

The Pour Point Process Analyser by BARTEC BENKE is an on-stream Analyser designed for closed-loop process control. It correctly measures pour point determinations for diesel, lube and heating oil production. The pour point is the temperature below which a liquid sample will not flow.

After thorough flushing the measuring cell is filled with the product. The product is cooled after a programmable settle down time. During cooling the measuring cell is tilted in 3˚C steps (other steps can be programmed) by a step motor. As soon as the IR-detector detects a movement on the surface of the product (sensitivity programmable), the measuring cell is moved back to vertical position.

If no movement is detected when the cell has been tilted to horizontal position, the measuring cell is kept in that position for 5 seconds. If also then no movement is detected, the temperature of the previous tilting is stored as the pour point temperature. Then the measuring cell is heated up, flushed and the next cycle starts.

PRODUCT SPECIFICATIONS

Type

4.1

Procedure

DIN / ISO 3016 / ASTM D-97

Comparability

≤ ISO / ASTM

Reproducibility

≤ ISO / ASTM

Measuring cycle

Discontinuous 15-90 Min. (depending on POUR POINT temp.)

Ambient temperature

10 - 40°C

Measuring Range

Please specify

Sample at analyzer inlet

Clean and dry, according to standard method

Flow rate

20 - 40 l/h

Pressure

100 - 300 kPa

Inlet temperature

Normal 30°C, max. 50°C, min. 20°C above Pour-Point temperature

Outlet / vent

Open to atmosphere

Accessories

Cooling water

Closed loop chiller, temperature depends on the measuring range, pressure 100 - 300 kPa, flow 20 - 40 l/h

Instrument air

Humidity class 1 (dew point ≤ -30°C)

Pressure

400 - 600 kPa constant over-pressure

Consumption

Approx. 1.4 - 1.8 Nm3 / per purge

Power supply / Electrical data

Analyzer Power supply

Please specify

Cooling water system

Please specify

Power consumption

Analyzer approx. 400W / chiller approx. 1600W

Input and output signals

2 x 4–20 mA, 200 W in
2 x 4–20 mA, 800 W out

General alarm

Potential – free contact (NO/NC)

Ready contact

Potential – free contact (NO/NC)