Flash Point Process Analyser

Flash Point Process Analyser Click here to enlarge image
Click here for Product Data Sheet

PRODUCT DESCRIPTION

The Flash Point Process Analyser (FPA-4) by BARTEC BENKE is a continuously measuring analyser suited to measure the flash point of liquid phase. It correlates with laboratory measurements. The FPA-4 operates on-line. It serves to monitor/maintain product quality for the in-spec production of mineral oil products.

The flash point temperature is defined as the lowest temperature at which application of an ignition source causes the vapour of a specimen of the sample to ignite under specified conditions of test. The sample flow continuously through the measuring cell of the FPA-4 and it is heated to the FP temperature and controlled by the FP detector.

The special features of the FPA-4 are continuous measurement, multi-stream capability, integrated failure diagnosis and self monitoring. Communication interfaces available include Modbus/RTU, Modbus/TCP, Remote Access via Modem, ISDN, LAN and VPN.

PRODUCT SPECIFICATIONS

Technical data

Method

ASTM D 56, ASTM D 93
DIN EN ISO 2719, DIN EN ISO 13736
IP 34, IP 170
DIN 51755

Measuring range

within 30 to 120 °C (86 to 248 °F) or within 80 to 160 °C (176 to 320 °F)
max. FP 160 °C (320 °F)

Repeatability

≤ ASTM

Reproducibility

≤ ASTM

Measuring cycle

continuously operated

Product streams

3 x sample, 1 x validation (additional hardware required)

Electrical data

Nominal voltage

AC 230 V ± 10 %, 1 phase; 50 Hz other rating on request

Maximum power consumption

approx. 700 W

Protection class

IP 54

Ambient conditions

Ambient temperature

operation 5 to 40 °C (41 to 104 °F)

Ambient humidity

operation 5 to 80 % relative humidity, non-corrosive

Sample

Quality

clean and dry; liquid (≤ 30 cSt)

Consumption

2 to 3 l/h (at sample inlet)

Pressure at inlet

1.5 to 3 bar

Temperature at inlet

min. 15 K below expected FP

Outlet

open to atmosphere

Utilities

Instrument air Consumption

min. 1.4 Nm³ per flushing cycle during start-up
~ 0.8 Nm3/h in normal operating mode only for leak compensation

Pressure at inlet

2 to 5 bar

Quality

dew point ≤ -40 °C (-40 °F)
humidity class 2 or better according to ISO8573.1

Signal outputs and inputs

Analog outputs

flash point
selectable

Digital outputs

sum alarm, measurement valid, see options

Digital inputs

reset, see options

Electrical data of signal outputs and inputs

Analog outputs

4 to 20 mA 800 Ω out;
active isolated on request

Digital outputs

DC 24 V; max. 0.5 A

Digital inputs

high DC 15 to 28 V
low DC 0 to 4 V

Auxiliary power supply output

DC 24 V; max. 0.8 A

Control unit

Central control unit

Industrial PC

Operating system

Windows XP®

Control software

PACS

User interfaces

Display

TFT display with touch function 800 x 600 pixels

Keyboard

virtual keyboard, controlled via TFT display

Connections

Pipe fittings

Swagelok® 6 mm/12 mm/18 mm other fittings on request

Vent/Slop

open to atmosphere

Weight and dimensions

Weight

approx. 300 kg

Dimensions (W x H x D)

approx. 1140 x 1900 x 710 mm

Optional signal outputs and inputs

Digital outputs

identification of a validation cycle
identification of a stream
warning/low-priority error
valve switching
calibration ID
regeneration ID

Digital inputs

stream selection
enable/disable automatic stream switching
request for a validation cycle
request calibration
request regeneration

MODBUS interface

MODBUS/RTU via RS485 or RS422 or fiber optic cable
MODBUS/TCP via fiber optic cable

Remote access

via modem, ISDN, Ethernet via fiber optical or VPN