Freezing Point Process Analyser

Freezing Point Process Analyser Click here to enlarge image
Click here for Product Data Sheet

PRODUCT DESCRIPTION

The Freezing Point Process Analyser (FRP) by BARTEC BENKE is an analyser for the fully automatic determination of the freezing point of liquid, transparent mineral oil products. The FRP operates on-line and is capable of handling up to two separate streams and a validation input. It serves to monitor/maintain product quality for the in-spec production of aviation turbine fuels and aviation gasoline.

The product sample is cooled down under defined conditions and its turbidity is observed. As soon as the crystallization point is reached, the point at which solid hydrocarbon crystals are detected, the cooling down of the sample will be stopped. The freezing point of the product sample is the temperature at which the solid hydrocarbon crystals, formed on cooling, completely disappear when the temperature of the fuel is allowed to rise.

The complete analysis procedure is controlled, monitored and visualised by the PACS (Process Analyser Control System) software. It offers a user interface for local analyser operation. The cooling performance of the analyser is achieved by means of peltier elements that are fed with the coolant of a chiller, which is an integral part of the FRP.

PRODUCT SPECIFICATIONS

Analyzer Type

FRP–4

Method

DIN ISO 3013 , ASTM D-2386, IP 16, ASTM D-5901

Comparability

≤ DIN EN / ASTM

Reproducibility

≤ DIN EN / ASTM

Measuring Cycle

approx. 8..20 minutes (depending on measuring range)

Environmental Temperature

5 .. 40°C

Environmental Humidity

Max. 70%, non – corrosive

Limit of Range

- 70°C (optional – 80°C)
the lowest detectable Freezing Point temperature depends on the actual Cloud Point temperature which must be above the mentioned limit

Sample at analyzer inlet (Clean and free of water according to applicable norms)

Flow Rate

5 .. 30 l/h

Pressure

2 .. 3 bar

Inlet Temperature

+ 5°C ..15°C

Outlet / Vent

Open to atmosphere

Utilities

Power Supply

Analyzer: See Specification of purged electronic housing
Chiller: 400VAC/3phs (as default); or acc. client's specification

Purging Gas
(Drying the EEx d enclosure)

Instrument Air or Nitrogen (consumption approx. 12 Nl / h)

Instrument Air (inlet)

See Specification of purged electronic housing

Electrical Signals

Output signals

1 x 4-20 mA, 800Ω out (standard), additional outputs on request

System alarm

Digital output / potential-free

Ready contact

Digital output / potential-free