Freezing Point Process Analyser
|
PRODUCT DESCRIPTION
The Freezing Point Process Analyser (FRP) by BARTEC BENKE is an analyser for the fully automatic determination of the freezing point of liquid, transparent mineral oil products. The FRP operates on-line and is capable of handling up to two separate streams and a validation input. It serves to monitor/maintain product quality for the in-spec production of aviation turbine fuels and aviation gasoline.
The product sample is cooled down under defined conditions and its turbidity is observed. As soon as the crystallization point is reached, the point at which solid hydrocarbon crystals are detected, the cooling down of the sample will be stopped. The freezing point of the product sample is the temperature at which the solid hydrocarbon crystals, formed on cooling, completely disappear when the temperature of the fuel is allowed to rise.
The complete analysis procedure is controlled, monitored and visualised by the PACS (Process Analyser Control System) software. It offers a user interface for local analyser operation. The cooling performance of the analyser is achieved by means of peltier elements that are fed with the coolant of a chiller, which is an integral part of the FRP.
|
PRODUCT SPECIFICATIONS
|
Analyzer Type
|
FRP–4
|
Method
|
DIN ISO 3013 , ASTM D-2386, IP 16, ASTM D-5901
|
Comparability
|
≤ DIN EN / ASTM
|
Reproducibility
|
≤ DIN EN / ASTM
|
Measuring Cycle
|
approx. 8..20 minutes (depending on measuring range)
|
Environmental Temperature
|
5 .. 40°C
|
Environmental Humidity
|
Max. 70%, non – corrosive
|
Limit of Range
|
- 70°C (optional – 80°C)
the lowest detectable Freezing Point temperature depends on the actual Cloud Point temperature which must be above the mentioned limit
|
Sample at analyzer inlet (Clean and free of water according to applicable norms)
|
|
Flow Rate
|
5 .. 30 l/h
|
Pressure
|
2 .. 3 bar
|
Inlet Temperature
|
+ 5°C ..15°C
|
Outlet / Vent
|
Open to atmosphere
|
Utilities
|
|
Power Supply
|
Analyzer: See Specification of purged electronic housing
Chiller: 400VAC/3phs (as default); or acc. client's specification
|
Purging Gas (Drying the EEx d enclosure)
|
Instrument Air or Nitrogen (consumption approx. 12 Nl / h)
|
Instrument Air (inlet)
|
See Specification of purged electronic housing
|
Electrical Signals
|
|
Output signals
|
1 x 4-20 mA, 800Ω out (standard), additional outputs on request
|
System alarm
|
Digital output / potential-free
|
Ready contact
|
Digital output / potential-free
|
|
|