FH62 C14 Continuous Ambient Particulate Monitor

FH62 C14 Continuous Ambient Particulate Monitor Click here to enlarge image
Click here for Product Data Sheet

PRODUCT DESCRIPTION

The FH62 C14 Continuous Ambient Particulate Monitor by Thermo Fisher Scientific measures the mass concentration of suspended particulate matter (PM) (TSP, PM-10, PM-2.5, PM-C and PM-1) by use of beta attenuation and in addition radon gas activity measurement eliminates interference of natural airborne radioactivity.

The FH62 C14 particulate sample collection area is located between both the C14 source and the proportional detector. While ambient particulate matter is being deposited onto a filter tape sample spot the dynamic filter loading is measured continuously by the attenuation of the C14 source beta rays and as a result a true continuous “real-time’ measurement of airborne particulate is provided.

This Ambient Particulate Monitor has user selectable reporting of mass concentration based on standard or actual flow rate and it is insensitive to vibration and diurnal temperatures (cycle of daily temperature changes).

PRODUCT SPECIFICATIONS

Measuring Principle

Continuous & simultaneous particulate collection coupled with beta ray attenuation

Source

Carbonium-14 (C14), <3.7 MBq (<100μCi)

Ranges

0 to 5,000 μg/m3 or 0 to 10,000 μg/m3

Minimum Detection Limit

<1 μg/m3 (24-hour average); <4 μg/m3 (1-hour average)

Precision of Two Monitors

± 2 μg/m3 (24-hour)

Resolution

± 1 μg/m3 (instantaneous)

Correlation Coefficient

R > 0.98

Measurement Cycle

Single filter spot in position for 24 hours (default); user selectable 30-minutes to 24 hours

Data Averages values displayed

Each full 1/2, and 24 hour values automatically stored; each full 1/2, 1, 3 and 24 hour

Air Flow Rate

1 m3/h (16.67 lpm) measured across an internal subsonic orifice; user selectable from 0 to 20 lpm

Output

Serial interface RS 232 Analog output: 4-20mA or 0-10V output of concentration (μg/m3)

Operating Temperature

-22 to 1400F (-30 to 600C)

Power Supply

Instrument: 100-240V, 50/60Hz, 330W max., 15W without pump or heater
Pump: 100-110/100-120V, 50/60Hz or 220/240V, 50/60Hz, 100W

Dimensions

Instrument: 19” (W) x 12.25” (H) x 13” (D) / 483mm (W) x 311mm (H) x 330mm (D)
Pump: 8.25” (W) x 8.75” (H) x 4.25” (D) / 210mm (W) x 222mm (H) x 108mm (D)

Weight

Instrument: 50 lbs (22.5 kg) / Pump: 18 lbs (8.16 kg)