Partisol-Plus 2025/2025-D Sequential Air Sampler
Regulatory Designations
- RFPS-0509-176: US EPA PM-Coarse Reference Sampler
- RFPS-0498-118: US EPA PM-2.5 Reference Sampler
- RFPS-1298-127: US EPA PM-10 Reference Sampler
- EN12341: Candidate equivalent PM-10 manual sampling method for European Norm EN12341.
Safety/Electrical Designations
Meets the following safety and electrical designations:
- CE: EN550011 Group 1, Class B (Emissions); EN55082-1 (Immunity); EN61010-1 (Safety)
- ETL: UL- and CSA-equivalent approval
Filter Holders and Media
Accommodates 47 mm filters. For USEPA PM-2.5 sampling, regulations only allow the use of Teflon® 2 μm pore size membrane filter. For PM-10 reference method sampling, the US EPA permits the use of Pallflex TX-40, quartz filter and Teflon® materials. Optional cassette separators may be used to eliminate passive volatilization losses or cross-contamination for speciation work.
Sample Flow Control, Reporting, and Audit/Calibration
- ActivolTM flow control system uses a mass flow controller and the measured ambient temperature and pressure to maintain a constant volumetric flow rate. the unit can control at rates ranging from 5 to 18 volumetric l/min.
- The sampler displays the current volumetric flow rate (l/min). For each filter exposed, the unit stores the total volume in volumetric or standard m3.
- Direct software support for single-point and multi-point audit/calibrations of the volumetric flow rate using a volumetric flow meter.
- Software support for the use of the StreamlineTM FTS Flow Transfer Standard for flow audits and calibrations.
Internal Data Storage
Interval Data (every 5 minutes), Input Data (stored at user-defined interval) and Filter Data (one record for each filter). This stored information may be viewed on the display and retrieved through the RS232 interface. The sampler has the following capacities:
- 16 days of Interval Data (stored every 5 minutes)
- 32 days of Input Data (stored every 30 minutes, default)
- 50 filter records
Data Input and Output
- Keypad/display for data retrieval and user programming.
- RS232 interface for data retrieval and remote operation. Supports advanced bi-directional AK Protocol.
- 3 user-defined analog outputs (0-5 Vdc) for data logging.
- 2 user-defined TTL alarm outputs.
- Wind vane/anemometer connection with 24 Vdc power output and two 2 0-5 Vdc inputs for wind speed and direction.
- 3 averaged analog inputs (0-5 Vdc) with conversion to engineering units (in addition to wind speed and direction above).
- RS485 interface for connection with other Thermo devices.
- Automatic calibration of analog inputs and output channels.
Meteorological Features
- The sampler always stores the following meteorological information in each record of Input Data and Filter Data: average ambient temperature, ambient pressure, average ambient relative humidity
- With an optional wind vane/anemometer connected to the unit, the following data are also stored in each record of Input Data and Filter Data: Average wind speed, wind velocity (vector averaged), average wind direction (vector averaged)
- In the case of Input Data, these values are averaged over the data storage interval, which is user-selectable (default 30 minutes). Data are sampled once per second.
- In the case of Filter Data, these values are averaged over the periods during which the hardware sampled through a given filter. Data are sampled once per second.
Operating Range, Dimensions and Power Requirements
- Temperature: -30 to +50 0C. Temperatures down to -40 °C require additional optional hardware.
- Enclosure: 27.2” (69.1 cm) W x 15.8” (40.2 cm) D x 26.3 (67.3 cm) H Height is 31” (78.8 cm) with the top cover, and 35.3” (89.5 cm) with the inlet connector.
- Weight: 115 lb (52 kg).
- Stand: Top Section: 30.8” (78.2 cm) W x 27.1” (68.8 cm) H x 14.0” (35.6 cm) D.
- Footprint: 42.0” (106.7 cm) W x 18.1” (46.0 cm) D.
- Power Requirements: 3 A @ 120 VAC, 1.5 A @ 240 VAC.