Lear Siegler Australasia

CEMS / Stack EmissionsCEMS / Stack EmissionsCEMS / Stack EmissionsCEMS / Stack Emissions

Beta Gauge F-904 Extractive Beta Gauge Particulate Monitor

Beta Gauge F-904 Extractive Beta Gauge Particulate Monitor Click here to enlarge image
Click here for Product Data Sheet

PRODUCT DESCRIPTION

The Beta Gauge F-904 Extractive Beta Gauge Particulate Monitor by VEREWA determines particulate concentration by measuring the amount of radiation a sample absorbs when exposed to a radioactive source. In general, the more energy absorbed, the greater the particulate concentration. The Model F904 is ideal for small diameter stack monitoring and there is a transportable version for mobile applications.

The typical applications that are suitable for the Model F-904 include coal and oil fired power plants, waste incinerators, waste water sludge incinerators, dust monitoring after wet scrubbers and heavy metal analysis. The instrument consists of five main modules. The sample probe in either stainless steel or titanium and these probes are suitable for either direct or diluted sample extraction and are heated. The sample is then passed through a heated sample line and is directed onto a filter tape. Once the gas passes through the filter tape it is routed to a downstream cooler to extract water and ensure a dry sample. A carbon vane rotary pump and mass flow controller ensure that the sample stream is pulled through these components at a flow rate of 3 cubic meters per hour.

The functions of the Model F-904 Extractive Beta Gauge Particulate Monitor are controlled by a powerful on-board microprocessor which calculates the particulate concentration value from the gas volume and zero/final radiation absorption differential.

PRODUCT SPECIFICATIONS

Ranges

selectable between 0-1 and 0-2000 mg/Nm3

Certified Ranges

0-5 bis 0-225mg/Nm3

Lower Detectable Limit

<0.3mg/Nm3

Total Error

<±5% F.S.

Zero Drift

automatic zero control

Span Drift

<1% F.S. / Week

MTBF

>95 % availability

Startup Time

30 min

Adjustment Time

2 min

Power Supply

230 V /50 Hz, 110V/60 Hz +10/-15 %, 5 kVA

Pressurized Air

6 bar, Instrument Air

Temperature Range

00C to +500C (32 to 1220F)

Signal Output

4-20 mA, Status Signals

Source

1 or 2 C-14-Flat Top Sources

Half Lifetime

5.730 years

Total Activity

<12,5 μCi /source

Detector

1 or 2 Geiger-Müller-Counter-Tubes

Filter Material

Glass-Fiber-Filter 99.95% >0.3mm, 150 m, 43 mm (1.7'') advance

Filter Adaptor

25 - 800C (77 to 1400F), temp. controlled (pre-conditioning of tape) adjustable

Sample Flow Rate

max. 3m3/h, user adjustable

Dilution Ratio

1:1 to 1:9

Cycle Time

user selectable, min. 5 minutes

V/R Operation (Forward/Backward)

max. 8 cycles (2 h average @ 15 min. cycle time)

Dimensions (H x W x D)

1600 x 800 x 800 mm (62 x 31 x 31'')

Weight

250 kg (550 lbs.)

Cabinet

Steel Cabinet with Front Door

Protection

IP 55 (NEMA-type)

Color

RAL 7032 (grey)

Sampling System

acc. VDI 2066, nozzle-Ø 5-12 mm (0.2-0.5''),
probe heated, isokinetic sampling, dilution, backflush, shut-off, material 1.4571 (SS) or titanium

Pump

Vacuum Rotary Pump, 3m3/h, 60% vacuum, w/ flow control

Sample gas cooler

Temperature +40C (390F), automatic condensate drain